全国服务热线
021-64195798CL-3000紫外交联仪(开门型)为CL-1000紫外交联仪升级替代款,利用中波302nm紫外线提供均匀强度的UV照射。 内部尺寸 254x127x305mm ,外部尺寸 396x216x348mm
INTRODUCTION
CL-3000紫外交联仪(开门型)为CL-1000紫外交联仪升级替代款,CL-3000紫外交联仪利用短波254nm、中波302nm、长波365nm紫外线提供均匀强度的UV照射。 CL-3000紫外交联仪内部尺寸 254x127x305mm ,外部尺寸 396x216x348mm
CL-3000紫外交联仪
占地面积小,台式使用
紫外强度和曝光时间可预设或手动调节
预曝光提供120,000微焦耳或五分钟的曝光
微处理器测量和控制紫外输出,可设置达999.900mj/cm2的高UV能量
5个302nm的中波紫外灯管提供均匀的紫外线
紫外安全门锁设计
大型LED显示屏和触摸屏,显示时间或能量设置
门上的窗口允许观察过程,同时阻挡紫外线辐射
短波UV型可选择门(CL-1000)或抽屉(CX-2000)样式
内置UV传感器在UVP工厂进行校准,以确保准确性和可靠性
与烘烤相比,将核酸连接到膜上的交联需要几秒钟
显示:LED
控制:触摸板
波长:302nm
紫外线曝光输出:120,000微焦耳或5分钟(可调);紫外能量设置可达999,900微焦/平方厘米
瓦特:8瓦
紫外光源:
CL-3000 254nm
CL-3000M 302nm
CL-3000L 365nm
CX-2000 254nm
CL-3000紫外交联仪的应用
核酸的膜固定
EB染色的DNA缺刻
检测recA 功能
利用胸腺嘧啶二聚体抑制切割进行基因作图
紫外灭菌
PCR污染消减等
CL-3000紫外交联仪的型号选择
型号 | 115V | 230V | 波长 | 可替换灯管 |
CL-3000(CL-1000) | 95-0174-01 | 95-0174-02 | 254nm | 34-0007-01 |
CL-3000L(CL-1000L) | 95-0228-01 | 95-0228-02 | 365nm | 34-0006-01 |
CL-3000M(CL-1000M) | 95-0230-01 | 95-0230-02 | 302nm | 34-0042-01 |
CX-2000 | 95-0339-01 | 95-0339-02 | 254nm | 34-0007-01 |
关于老款UVP CL-1000紫外交联剂的说明:
老款UVP CL-1000紫外交联剂已于2020年8月1日停产。自停产之日起,UVP紫外交联剂CL-1000,CL-1000M和CL-1000L的服务和备件将提供5年(2025年8月1日)。 。
CL-3000 General Applications
UV induced mutations
Membrane crosslinking
Biomolecular crosslinking
UV dose calibrated to a NIST traceable standard
Product Details
Built-in radiometer calibrated to a NIST traceable standard enables precise and accurate dosing to obviate end-user calibration
Safety interlocks to prevent users from accidental UV exposure
Capable of delivering up 10 J/cm2 of UVA, UVB, or UVC
Extensive publication record using UVP Crosslinkers (here)
Highly-uniform surface illumination
Compact-size to accommodate limited lab space
hortwave UV (UVC) as a Disinfectant
Ultraviolet light is considered an ancillary disinfectant to chemical disinfectants in the healthcare environment. Our high-precision CL-3000 crosslinker is capable of producing a cumulative dosage of 1 J/cm2 , this is considered well beyond the requirement needed for disinfection. The CL-3000 provides nearuniform UVC light with a ~10% coefficient of variation and less than 2 mJ/cm2 difference between the highest and lowest reading. Most importantly, the CL-3000 has real-time monitoring of UVC dosage to assure reproducible dosing irrespective of space and time.
Immobilization of Nucleic Acids to Membranes
The UVP Crosslinker is a microprocessor controlled UV irradiation system dedicated to nucleic acid linking to membranes for Southern, Northern, Dot and Slot Blot applications. It can also be used for UV disinfection and for elimination of PCR contaminations.
Microprocessor Controlled Reproducibility
The programmable microprocessor constantly monitors the UV light emission. Irradiation stops exactly when the programmed energy is achieved. Thus, the effect of decreasing UV intensity due to bulb aging is compensated for.
Durability
The UVP Crosslinker combines the latest UV technology with high quality manufacturing: aluminum UV exposure chamber, protective quartz disk on the UV sensor cell, and a highly resistant and easy-to-clean keypad and housing.
Ease of use
The large display provides a series of predefined methods, making the UVP Crosslinker an easy to use yet powerful instrument for immobilization of nucleic acids to membranes. The programmed data are shown on the LED display.
UVLink 1000 Crosslinker/Legacy UVP Crosslinkers
Legacy UVP Crosslinker CL-1000 were discontinued August 1st, 2020. Service and spare parts for UVP Crosslinker CL-1000, CL-1000M, and CL-1000L will be available and offered for 5 years from discontinuation date (August 1st 2025).
Technical Specifications | CL-3000 | CL-3000M | CL-3000L | |
Wavelengths | 254 nm | 302 nm | 365 nm | |
Bulbs | 6 x 8 Watt | |||
Energy | 0000.1 - 9999.9 mJ/cm² (0 - 10 J/cm²) | |||
Time | 000:01 - 999:59 (>300 J/cm²) | |||
Temperature | 15⁰C - 35⁰C | |||
Humidity | 70% Non-Condensing | |||
Altitude | up to 3,000M (9,842 ft) | |||
Sound Level | ≤ 50 dba | |||
Housing Surface Temp | ≤ 30⁰C | |||
Startup Time | < 1 sec | |||
External Dim (L x W x H) | 41 cm x 40 cm x 26.5 cm | |||
Internal Dim (L x W x H) | 35 cm x 27 cm x 16 cm | |||
Weight | 6.8 Kg: 15 lb | |||
Operating Power | 100 - 115VAC & 230VAC 50/60Hz | |||
Certifications | CE, RoHS (CSA In Process) |
Order number | Item | |||
---|---|---|---|---|
115 V | 230 V | Description | Wavelength | |
849-95-0615-01 | 849-95-0615-02 | UVP Crosslinker CL-3000 | 254 nm | |
849-95-0615-03 | 849-95-0615-04 | UVP Crosslinker CL-3000M | 302 nm | |
849-95-0615-05 | 849-95-0615-06 | UVP Crosslinker CL-3000L | 365 nm | |
34-0007-01 (US) 849-057-007 (EU) | Replacement UV tube, 8 Watt, 254 nm | |||
34-0042-01 (US) | Replacement UV tube, 8 Watt, 302 nm | |||
34-0006-01 (US) | Replacement UV tube, 8 Watt, 365 nm |